Signal-correction errors in the EasyOne Pro LAB multiple-breath washout device significantly impact outcomes in children and adults

Marc Alexander Oestreich, Florian Wyler, Christian Buess, Blaise Etter, Kathryn A. Ramsey, Philipp Latzin

Research output: Contribution to journalArticlepeer-review

1 Citation (Scopus)

Abstract

Multiple-breath washout (MBW) is an established technique to assess functional residual capacity (FRC) and ventilation inhomogeneity in the lung. Indirect calculation of nitrogen concentration requires accurate measurement of gas concentrations. To investigate the accuracy of the CO2 concentration and molar mass (MM) values used for the indirect calculation of nitrogen concentration in a commercial MBW device [EasyOne Pro LAB (EOPL), ndd Medizintechnik AG, Switzerland] and its impact on outcomes. We used high-precision gas mixtures to evaluate CO2 and MM sensor output in vivo and in vitro. We developed updated algorithms to correct observed errors and assessed the impact on MBW outcomes and FRC measurement accuracy compared with body plethysmography. The respiratory exchange ratio (RER)-based adjustment of the measured CO2 signal used in the EOPL led to an overestimated CO2 signal (range -0.1% to 1.0%). In addition, an uncorrected dependence on humidity was identified. These combined effects resulted in an overestimation of expired nitrogen concentrations (range -0.7% to 2.6%), and consequently MBW outcomes. Corrected algorithms reduced the mean (SD) cumulative expired volume by 15.8% (9.7%), FRC by 6.6% (3.0%), and lung clearance index by 9.9% (7.6%). Differences in FRC between the EOPL and body plethysmography further increased. Inadequate signal correction causes RER- and humidity-dependent expired nitrogen concentration errors and overestimation of test outcomes. Updated algorithms reduce average signal error, however, RER values far from the population average still cause measurement errors. Despite improved signal accuracy, the updated algorithm increased the difference in FRC between the EOPL and body plethysmography.

Original languageEnglish
Pages (from-to)460-471
Number of pages12
JournalJournal of Applied Physiology
Volume136
Issue number3
DOIs
Publication statusPublished - Mar 2024
Externally publishedYes

Fingerprint

Dive into the research topics of 'Signal-correction errors in the EasyOne Pro LAB multiple-breath washout device significantly impact outcomes in children and adults'. Together they form a unique fingerprint.

Cite this