Reproducibility of the creamatocrit technique for the measurement of fat content in human milk

Research output: Contribution to journalArticlepeer-review

4 Citations (Scopus)

Abstract

Quantification of human milk (HM) fat is important for determining the energy intake of infants. The simplest and most rapid method is the creamatocrit method. However, the reliability of the creamatocrit has not been comprehensively investigated. The aims of this study were to test the inter- and-intra-rater reliability of: 1) HM sampling after hand- or-machine mixing methods and 2) HM fat measurement by the creamatocrit method. Inter-and-intra rater HM sampling after hand- or-machine mixing methods had high intraclass correlation coefficient (>0.91). Inter-rater reliability of measurement of HM with low fat (<2%) resulted in high variability (median coefficient of variations (CVs) > 15%). Intra- and inter-rater reliability of measurement of HM with higher fat (>3.5%) had low variability (median CVs < 10%). As the greatest variation in the creamatocrit method occurred during the measurement of HM samples with low fat, duplicate readings are necessary to reduce discrepancies in every HM fat determination.

Original languageEnglish
Article number129708
JournalFood Chemistry
Volume356
DOIs
Publication statusPublished - 15 Sept 2021

Fingerprint

Dive into the research topics of 'Reproducibility of the creamatocrit technique for the measurement of fat content in human milk'. Together they form a unique fingerprint.

Cite this