Paediatric radiation dose measurements undergoing X-ray examination in Tasmania

Vipul Agarwal, Stephen Newbery

Research output: Contribution to journalArticlepeer-review

Abstract

A pilot study was carried out involving six radiology centres across the state of Tasmania, Australia, with two main goals. First, to compare the choice of exposure factors used by each centre for four common paediatric projections of a 'standard' infant, 5-, 10- and 15-year-old patients. Second, to measure dose area product (DAP) at each radiology centre using the factors obtained and compare it against published diagnostic reference levels (DRLs). Significant variation was observed between the six radiology centres studied which has been attributed to the variability in exposure factors provided by the centres. Notably, the measured DAP values in the case of infant and 5-year-old patients were observed to be significantly lower than the published reference DAP DRLs. While significantly higher exposure factors have been observed for 15-year-old patients as illustrated in measured DAP values.

Original languageEnglish
Article numberncw143
Pages (from-to)382-386
Number of pages5
JournalRadiation Protection Dosimetry
Volume174
Issue number3
DOIs
Publication statusPublished - 2017

Fingerprint

Dive into the research topics of 'Paediatric radiation dose measurements undergoing X-ray examination in Tasmania'. Together they form a unique fingerprint.