Microstructural features in fractured high nitrogen stainless steel hip prostheses: a retrieval study of polished, tapered femoral stems

E. Swarts, A. Kop, N. Jones, C. Keogh, S. Miller, Piers Yates

    Research output: Contribution to journalArticlepeer-review

    6 Citations (Scopus)

    Abstract

    Six fractured high nitrogen stainless steel (HNSS), polished, tapered hip stems have been reviewed for analysis. Clinical data suggests that poor proximal support (loosening), varus positioning, and canal morphology are implicated in the failure of these stems. Metallurgical assessment reveals a number of microstructural deficiencies contributing to failure. Of significance is poor grain homogeneity and larger grain size near the surface of the fractured stems. It appears some polished, tapered HNSS stems are at risk when combinations of these clinical and metallurgical features occur. As the control of grain size, microstructural uniformity, and inclusion content are critical for optimum mechanical performance, it is important the manufacturers review processing methods to further minimize the risk of component failure. © 2007 Wiley Periodicals, Inc. J Biomed Mater Res 84A: 753–760, 2008
    Original languageEnglish
    Pages (from-to)753-760
    JournalJournal of Biomedical Materials Research Part A
    Volume84A
    Issue number3
    DOIs
    Publication statusPublished - 2008

    Fingerprint

    Dive into the research topics of 'Microstructural features in fractured high nitrogen stainless steel hip prostheses: a retrieval study of polished, tapered femoral stems'. Together they form a unique fingerprint.

    Cite this