Local T1-T2 distribution measurements in porous media

Sarah Vashaee, Ming Li, Ben Newling, Bryce MacMillan, Florin Marica, Hyung Kwak, J Gao, AM Al-Harbi, Bruce Balcom

Research output: Contribution to journalArticle

4 Citations (Scopus)

Abstract

A novel slice-selective T1-T2 measurement is proposed to measure spatially resolved T1-T2 distributions. An adiabatic inversion pulse is employed for slice-selection. The slice-selective pulse is able to select a quasi-rectangular slice, on the order of 1 mm, at an arbitrary position within the sample. The method does not employ conventional selective excitation in which selective excitation is often accomplished by rotation of the longitudinal magnetization in the slice of interest into the transverse plane, but rather a subtraction based on CPMG data acquired with and without adiabatic inversion slice selection. T1 weighting is introduced during recovery from the inversion associated with slice selection. The local T1-T2 distributions measured are of similar quality to bulk T1-T2 measurements. The new method can be employed to characterize oil–water mixtures and other fluids in porous media. The method is beneficial when a coarse spatial distribution of the components is of interest. © 2018 Elsevier Inc.
Original languageEnglish
Pages (from-to)113-122
JournalJournal of Magnetic Resonance
Volume287
DOIs
Publication statusPublished - Feb 2018
Externally publishedYes

    Fingerprint

Cite this

Vashaee, S., Li, M., Newling, B., MacMillan, B., Marica, F., Kwak, H., ... Balcom, B. (2018). Local T1-T2 distribution measurements in porous media. Journal of Magnetic Resonance, 287, 113-122. https://doi.org/10.1016/j.jmr.2018.01.001