Electronic portal imaging vs kilovoltage imaging in fiducial marker image-guided radiotherapy for prostate cancer: An analysis of set-up uncertainties

Suki Gill, J. Thomas, C. Fox, T. Kron, A. Thompson, S. Chander, S. Williams, K. H. Tai, G. Duchesne, F. Foroudi

Research output: Contribution to journalArticlepeer-review

10 Citations (Scopus)

Abstract

Objectives: The purpose of this study was to compare interfraction prostate displacement data between electronic portal imaging (EPI) and kilovoltage imaging (KVI) treatment units and discuss the impact of any difference on margin calculations for prostate cancer image-guided radiotherapy (IGRT). Methods: Prostate interfraction displacement data was collected prospectively for the first 4 fractions in 333 patients treated with IGRT with daily pre-treatment EPI or KVI orthogonal imaging. Displacement was recorded in the anteroposterior (AP), left-right (LR) and superoinferior (SI) directions. The proportion of displacement <3 mm and the difference in median absolute displacements were calculated in all directions. Results: 1088 image pairs were analysed in total, 448 by EPI and 640 by KVI. There were 23% (95% confidence interval [CI] 18-28%) more displacements under 3mm for EPI than for KVI in the AP direction, 14% (95% CI 10-19%) more in the LR direction and 10% (95% CI 5-15%) more in the SI direction. The differences in absolute median displacement (KVI>EPI) were AP 1 mm, LR 1 mm and SI 0.5 mm. Wilcoxon rank-sum test showed that distributions were significantly different for all three dimensions (p<0.0001 for AP and LR and p=0.02 for SI). Conclusion: EPI has a statistically significant smaller set-up error distribution than KVI. We would expect that, because fiducial marker imaging is less clear for EPI, the clinical target volume to planning target volume margin would be greater when using IGRT; however, relying wholly on displacement data gives the opposite result. We postulate that this is owing to observer bias, which is not accounted for in margin calculation formulas.

Original languageEnglish
Pages (from-to)176-182
Number of pages7
JournalBritish Journal of Radiology
Volume85
Issue number1010
DOIs
Publication statusPublished - Feb 2012
Externally publishedYes

Fingerprint

Dive into the research topics of 'Electronic portal imaging vs kilovoltage imaging in fiducial marker image-guided radiotherapy for prostate cancer: An analysis of set-up uncertainties'. Together they form a unique fingerprint.

Cite this