N2 - The importance of appropriate equipment to manage the difficult airway has been highlighted by the publication of the Australian and New Zealand College of Anaesthetists (ANZCA) guidelines in 2012. We set out to audit compliance with these guidelines in all public and private sites providing general anaesthesia in metropolitan Perth. Public and private healthcare websites identified 39 sites of which 37 were studied. Institutional and ethics approval was obtained. A tick-box design audit tool, based on the ANZCA guidelines, was used to collect information regarding the dedicated difficult airway container (DDAC) at each site. As recommended in the guidelines, only equipment within the DDAC was considered. Further data about each site, including the number of theatre suites, satellite anaesthetic areas, use of capnography and categories of patients treated (adult, obstetric and paediatric) were collected. An adult DDAC was found at 92% of all sites, but none of the sites had all the essential equipment listed in the ANZCA guidelines. There was limited provision of adult difficult airway equipment within private sites compared to public, and less provision of paediatric difficult airway equipment across all sites treating paediatric patients in metropolitan Perth. Capnography was available in 76% of post anaesthesia care units and used regularly in 27%. Adherence to the ANZCA guidelines regarding the DDAC could be improved. Standardised equipment across a metropolitan region would be of value in the management of the difficult airway.
AB - The importance of appropriate equipment to manage the difficult airway has been highlighted by the publication of the Australian and New Zealand College of Anaesthetists (ANZCA) guidelines in 2012. We set out to audit compliance with these guidelines in all public and private sites providing general anaesthesia in metropolitan Perth. Public and private healthcare websites identified 39 sites of which 37 were studied. Institutional and ethics approval was obtained. A tick-box design audit tool, based on the ANZCA guidelines, was used to collect information regarding the dedicated difficult airway container (DDAC) at each site. As recommended in the guidelines, only equipment within the DDAC was considered. Further data about each site, including the number of theatre suites, satellite anaesthetic areas, use of capnography and categories of patients treated (adult, obstetric and paediatric) were collected. An adult DDAC was found at 92% of all sites, but none of the sites had all the essential equipment listed in the ANZCA guidelines. There was limited provision of adult difficult airway equipment within private sites compared to public, and less provision of paediatric difficult airway equipment across all sites treating paediatric patients in metropolitan Perth. Capnography was available in 76% of post anaesthesia care units and used regularly in 27%. Adherence to the ANZCA guidelines regarding the DDAC could be improved. Standardised equipment across a metropolitan region would be of value in the management of the difficult airway.
M3 - Article
SN - 0310-057X
VL - 42
SP - 657
EP - 664
JO - Anaesthesia and Intensive Care
JF - Anaesthesia and Intensive Care
IS - 5
ER -