TY - JOUR
T1 - Comprehensive Australasian multicentre dosimetric intercomparison: Issues, logistics and recommendations
AU - Ebert, Martin
AU - Harrison, K.M.
AU - Comes, D.
AU - Howlett, S.J.
AU - Joseph, D.J.
AU - Kron, T.
AU - Hamilton, C.S.
AU - Denham, J.W.
PY - 2009
Y1 - 2009
N2 - The present paper describes the logistics of the 2004–2008 Australasian Level III Dosimetry Intercomparison. Dosimetric intercomparisons (or ‘audits’) can be used in radiotherapy to evaluate the accuracy and quality of radiation delivery. An intercomparison was undertaken in New Zealand and Australia to evaluate the feasibility and logistics of ongoing dosimetric intercomparisons that evaluate all steps in the radiotherapy treatment process, known as a ‘Level III’ intercomparison. The study commenced in 2002 with the establishment of a study team, definition of the study protocol, acquisition of appropriate equipment and recruitment of participating radiotherapy centres. Measurements were undertaken between October 2004 and March 2008, and included collation of data on time, costs and logistics of the study. Forty independent Australian and New Zealand radiotherapy centres agreed to participate. Measurement visits were made to 37 of these centres. Data is presented on the costs of the study and the level of support required. The study involved the participation of 16 staff at the study centre who invested over 4000 hours in the study, and of over 200 professionals at participating centres. Recommendations are provided for future phantom-based intercomparisons. It is hoped that the present paper will be of benefit to any centres or groups contemplating similar activities by identifying the processes involved in establishing the study, the potential hazards and pitfalls, and expected resource requirements.
AB - The present paper describes the logistics of the 2004–2008 Australasian Level III Dosimetry Intercomparison. Dosimetric intercomparisons (or ‘audits’) can be used in radiotherapy to evaluate the accuracy and quality of radiation delivery. An intercomparison was undertaken in New Zealand and Australia to evaluate the feasibility and logistics of ongoing dosimetric intercomparisons that evaluate all steps in the radiotherapy treatment process, known as a ‘Level III’ intercomparison. The study commenced in 2002 with the establishment of a study team, definition of the study protocol, acquisition of appropriate equipment and recruitment of participating radiotherapy centres. Measurements were undertaken between October 2004 and March 2008, and included collation of data on time, costs and logistics of the study. Forty independent Australian and New Zealand radiotherapy centres agreed to participate. Measurement visits were made to 37 of these centres. Data is presented on the costs of the study and the level of support required. The study involved the participation of 16 staff at the study centre who invested over 4000 hours in the study, and of over 200 professionals at participating centres. Recommendations are provided for future phantom-based intercomparisons. It is hoped that the present paper will be of benefit to any centres or groups contemplating similar activities by identifying the processes involved in establishing the study, the potential hazards and pitfalls, and expected resource requirements.
U2 - 10.1111/j.1754-9485.2009.02047.x
DO - 10.1111/j.1754-9485.2009.02047.x
M3 - Article
C2 - 19453538
SN - 1754-9477
VL - 53
SP - 119
EP - 131
JO - Journal of Medical Imaging and Radiation Oncology
JF - Journal of Medical Imaging and Radiation Oncology
ER -