Comparison of the efficacy and safety of cuffed versus uncuffed endotracheal tubes for infants in the intensive care setting: a pilot, unblinded RCT

Rebecca E. Thomas, Simon Erickson, Bruce Hullett, Corrado Minutillo, Martyn Lethbridge, Shyan Vijayasekaran, Sachin Agrawal, Max K. Bulsara, Shripada C. Rao

Research output: Contribution to journalArticlepeer-review

5 Citations (Scopus)

Abstract

Objective To study effectiveness and safety of cuffed versus uncuffed endotracheal tubes (ETTs) in small infants in the intensive care unit (ICU).

Design Pilot RCT.

Setting Neonatal and paediatric ICUs of children's hospital in Western Australia.

Participants Seventy-six infants >= 35 weeks gestation and infants = 3 kg.

Interventions Patients randomly assigned to Microcuff cuffed or Portex uncuffed ETT.

Main outcomes measures Primary outcome was achievement of optimal ETT leak in target range (10%-20%). Secondary outcomes included: reintubations, ventilatory parameters, ventilatory complications, postextubation complications and long-term follow-up.

Results Success rate (achievement of mean leak in the range 10%-20%) was 13/42 (30.9%) in the cuffed ETT group and 6/34 (17.6%) in uncuffed ETT group (OR=2.09; 95% CI (0.71 to 6.08); p=0.28). Mean percentage time within target leak range in cuffed ETT group 28% (IQR: 9-42) versus 15% (IQR: 0-28) in uncuffed ETT group (p=0.01). There were less reintubations to optimise size in cuffed ETT group 0/40 versus 10/36 (p

Conclusions There was no difference in the primary outcome, though percentage time spent in optimal leak range was significantly higher in cuffed ETT group. Cuffed ETTs reduced reintubations to optimise ETT size and episodes of atelectasis. Cuffed ETTs may be a feasible alternative to uncuffed ETTs in this group of patients.

Original languageEnglish
Pages (from-to)F614-F620
Number of pages7
JournalADC Fetal & Neonatal
Volume106
Issue number6
DOIs
Publication statusPublished - 1 Nov 2021
Externally publishedYes