Cluster hole versus solid cup in total hip arthroplasty: A randomized control trial: a randomized control trial

    Research output: Contribution to journalArticlepeer-review

    7 Citations (Scopus)

    Abstract

    © 2014. Acetabular osteolysis has been linked to polyethylene debris that is generated in the hip migrating through screw holes in the acetabular component. Solid-backed acetabular components were designed to decrease this osteolysis. This prospective trial randomized 100 patients undergoing total hip arthroplasty to either a solid-backed or a cluster-hole acetabular component-all without screws. At 5. years post-surgery, 34.4% of all patients had osteolytic lesions that were visible on CT. There was no significant difference in either presence or volume of the osteolytic lesions, cup migration or functional outcomes (OHS) between the groups. There may no longer be a detriment to using cluster-hole cups instead of solid cups in all hips. This would then give the surgeon the option to use screws for stability as required.
    Original languageEnglish
    Pages (from-to)223-229
    Number of pages7
    JournalJournal of Arthroplasty
    Volume30
    Issue number2
    DOIs
    Publication statusPublished - 2015

    Fingerprint

    Dive into the research topics of 'Cluster hole versus solid cup in total hip arthroplasty: A randomized control trial: a randomized control trial'. Together they form a unique fingerprint.

    Cite this