Analysis of "Accuracy evaluation of five blood glucose monitoring systems: The North American comparator trial"

    Research output: Contribution to journalArticlepeer-review

    1 Citation (Scopus)

    Abstract

    In an article in Journal of Diabetes Science and Technology, Halldorsdottir and coauthors examined the accuracy of five blood glucose monitoring systems (BGMSs) in a study sponsored by the manufacturer of the BGMS CONTOUR NEXT EZ (EZ) and found that this BGMS was the most accurate one. However, their findings must be viewed critically given that one of the BGMSs (ACCU-CHEK Aviva) was not compared against the reference measurement specified by its manufacturer, thus making it likely that it performed suboptimally. Also, the accuracy of the glucose-oxidase-based ONE TOUCH Ultra2 and TRUEtrack BGMS is likely to have been underestimated because of the expected low oxygen level in the glycolysed blood samples used to test the performance of these BGMSs under hypoglycemic conditions. In conclusion, although this study shows that EZ is an accurate BGMS, comparisons between this and other BGMSs should be interpreted with caution. © Diabetes Technology Society.
    Original languageEnglish
    Pages (from-to)1305-1307
    JournalJournal of Diabetes Science and Technology
    Volume7
    Issue number5
    DOIs
    Publication statusPublished - 2013

    Fingerprint

    Dive into the research topics of 'Analysis of "Accuracy evaluation of five blood glucose monitoring systems: The North American comparator trial"'. Together they form a unique fingerprint.

    Cite this