An audit of the diagnostic accuracy of rotational thromboelastometry for the identification of hypofibrinogenaemia and thrombocytopenia during cardiopulmonary bypass

S. A. Matzelle, W. M. Weightman, N. M. Gibbs

Research output: Contribution to journalArticle

Abstract

We audited the diagnostic accuracy of ROTEM (R) (TEM Innovations, GmbH, Munich, Germany) measurements of hypofibrinogenaemia (fibrinogen 8 mm and PLTEM A10 values >= 35 mm, but would continue to use laboratory measurements for confirmation. We would not use FIBTEM A10

Original languageEnglish
Pages (from-to)620-626
Number of pages7
JournalAnaesthesia and Intensive Care
Volume46
Issue number6
DOIs
Publication statusPublished - Nov 2018
Externally publishedYes

Cite this

@article{63ca2fde92e945c7a4bdb48ecbd4e972,
title = "An audit of the diagnostic accuracy of rotational thromboelastometry for the identification of hypofibrinogenaemia and thrombocytopenia during cardiopulmonary bypass",
abstract = "We audited the diagnostic accuracy of ROTEM (R) (TEM Innovations, GmbH, Munich, Germany) measurements of hypofibrinogenaemia (fibrinogen 8 mm and PLTEM A10 values >= 35 mm, but would continue to use laboratory measurements for confirmation. We would not use FIBTEM A10",
keywords = "cardiopulmonary bypass, hypofibrinogenaemia, thrombocytopenia, thromboelastometry, CARDIAC-SURGERY, COAGULATION MANAGEMENT, GUIDELINES, TESTS, METAANALYSIS, ROTEM",
author = "Matzelle, {S. A.} and Weightman, {W. M.} and Gibbs, {N. M.}",
year = "2018",
month = "11",
doi = "10.1177/0310057X1804600614",
language = "English",
volume = "46",
pages = "620--626",
journal = "Anaesthesia Intensive Care",
issn = "0310-057X",
publisher = "Australian Society of Anaesthetists",
number = "6",

}

TY - JOUR

T1 - An audit of the diagnostic accuracy of rotational thromboelastometry for the identification of hypofibrinogenaemia and thrombocytopenia during cardiopulmonary bypass

AU - Matzelle, S. A.

AU - Weightman, W. M.

AU - Gibbs, N. M.

PY - 2018/11

Y1 - 2018/11

N2 - We audited the diagnostic accuracy of ROTEM (R) (TEM Innovations, GmbH, Munich, Germany) measurements of hypofibrinogenaemia (fibrinogen 8 mm and PLTEM A10 values >= 35 mm, but would continue to use laboratory measurements for confirmation. We would not use FIBTEM A10

AB - We audited the diagnostic accuracy of ROTEM (R) (TEM Innovations, GmbH, Munich, Germany) measurements of hypofibrinogenaemia (fibrinogen 8 mm and PLTEM A10 values >= 35 mm, but would continue to use laboratory measurements for confirmation. We would not use FIBTEM A10

KW - cardiopulmonary bypass

KW - hypofibrinogenaemia

KW - thrombocytopenia

KW - thromboelastometry

KW - CARDIAC-SURGERY

KW - COAGULATION MANAGEMENT

KW - GUIDELINES

KW - TESTS

KW - METAANALYSIS

KW - ROTEM

U2 - 10.1177/0310057X1804600614

DO - 10.1177/0310057X1804600614

M3 - Article

VL - 46

SP - 620

EP - 626

JO - Anaesthesia Intensive Care

JF - Anaesthesia Intensive Care

SN - 0310-057X

IS - 6

ER -