title = "An audit of the diagnostic accuracy of rotational thromboelastometry for the identification of hypofibrinogenaemia and thrombocytopenia during cardiopulmonary bypass",
abstract = "We audited the diagnostic accuracy of ROTEM (R) (TEM Innovations, GmbH, Munich, Germany) measurements of hypofibrinogenaemia (fibrinogen 8 mm and PLTEM A10 values >= 35 mm, but would continue to use laboratory measurements for confirmation. We would not use FIBTEM A10",
keywords = "cardiopulmonary bypass, hypofibrinogenaemia, thrombocytopenia, thromboelastometry, CARDIAC-SURGERY, COAGULATION MANAGEMENT, GUIDELINES, TESTS, METAANALYSIS, ROTEM",
author = "Matzelle, {S. A.} and Weightman, {W. M.} and Gibbs, {N. M.}",
year = "2018",
month = nov,
doi = "10.1177/0310057X1804600614",
language = "English",
volume = "46",
pages = "620--626",
journal = "Anaesthesia Intensive Care",
issn = "0310-057X",
publisher = "Australian Society of Anaesthetists",
number = "6",
}