Accuracy and reliability of a cosmed K4b(2) portable gas analysis system

R.J. Duffield, Brian Dawson, H. Pinnington, P. Wong

Research output: Contribution to journalArticlepeer-review

270 Citations (Scopus)

Abstract

The purpose of this study was to assess the validity and reliability of a Cosmed K4b 2 portabletelemetric gas analysis system. Twelve physically fit males performed a treadmill runningsession consisting of an easy 10 rain run, a hard 3 min run and a 1 min sprint (with restperiods of 10 rain separating each run), on four separate occasions. Sessions were identicalwith the exception of the apparatus used to measure VO 2. Du~_ng two (test-retest) sessionsa Cosmed K4b 2 portable gas analysis system was used; in another, a laboratory metaboliccart and, in one session, both systems were used to measure ~'O 2 simultaneously.Comparison of Cosmed K4b 2 and metabolic cart measurements in isolation revealedsignificantly (p
Original languageEnglish
Pages (from-to)11-22
JournalJournal of Science and Medicine in Sport
Volume7
Issue number1
Publication statusPublished - 2004