Abstract
To assess osteoarthritic changes in knee joints a radiography rig for acquisition of standardized radiographs of trabecular bone has been developed. The rig contains a steel frame on castors, a turntable, a cassette holder frame, calibration Plexiglas sheets, body supports and points. It is used to lock the patient in a standardized position. A film cassette holder frame was also developed to reduce scattering of X-rays, and consequently the amount of noise in the radiographs. Calibration Plexiglas sheets were mounted on ball bearing slides to obtain radiographs without a calibration pattern (suitable for the analysis of trabecular bone texture) and radiographs containing a calibration pattern (suitable for the measurement of leg alignment).
Original language | English |
---|---|
Pages (from-to) | 101-103 |
Journal | Acta Radiologica |
Volume | 43 |
Issue number | 2 |
DOIs | |
Publication status | Published - 2002 |